Error code
E19005
E19402
E19403
E19404
E19405
E19406
E19407
E19408
E19409
Pivot
movement
errors
E21xxx:
E21002
ORTHOPANTOMOGRAPH
Problem
Bus failing
ESD error ref
ESD error signal
ESD error proximity
ESD error bus
Slider disabled too large change
Slider disabled all channels
active
Proximity active after reset
Slider disabled proximity active
Route - waypoints on invalid
state
OP 3D
Solution
1. Check connections:
* J4007, J6001
If fail => replace user interface cable
2. Replace the upper cover of the
lower shelf (contains the UI panel
module)
3. Replace i4000.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
1. Check that there are no objects on
the UI panel.
Remove any objects.
2. Restart the device.
If error still exists => Replace
the upper cover of the lower shelf
(contains the UI panel module with
i6000).
1. Check that there are no objects on
the UI panel.
Remove any objects.
2. Restart the device.
If error still exist => Replace the
upper cover of the lower shelf
(contains the UI panel module with
i6000).
Software error. Take logs and boot the
device.
7 Troubleshooting
81