6 Troubleshooting

6.1 Check for malfunctions

Missing or damaged O-rings.
Malfunction and premature failure.
▶ Make sure that all O-rings are on the coupling and are undamaged.
Heating of the product.
Burns or product damage from overheating.
▶ Do not use the product if it is irregularly heated.
▶ The medical device is too hot while working:
Service the medical device.
▶ When the speed drops or is uneven:
Service the medical device.
▶ An O-ring is missing on the motor coupling:
Replace O-ring.
See also:
2 Instructions for use of motor
6.2 Troubleshooting

6.2.1 Replacing the O-rings

Hazard from improper care of the O-rings.
Functional problem or complete failure of the medical device.
▶ Do not use Vaseline or other grease or oil.
Note
The O-rings on the coupling may only be lubricated with a cotton ball wetted with
KaVo Spray.
▶ Press the O-ring between your fingers to form a loop.
▶ Push the O-ring to the front, and remove it.
▶ Insert new O-rings into the grooves.
Instructions for use SMARTmatic S80 K - 1.011.6771