Error code
E22203
E22204
E22205
E22206
E22207
E22208
E22209
E22210
E22211
E22212
E22213
Rotation
movement
errors
E23xxx:
E23002
E23202
ORTHOPANTOMOGRAPH
Problem
Calib ABS - opto hit invalid ADC
value
Calib ABS - opto hit ADC value
change too small
Calib ABS - opto hit too early
Calib ABS verify - too large error See E09018
Route - no valid route found
Goto ADC - Invalid start position Software error. Take logs and boot the
Goto ADC - Opto hit too fast
Goto ADC - Opto edge missing
Goto ADC - Edges end not found 1. Check the location of the resistive
Goto ADC - Adc value not
changing
Goto ADC - Adc value changing
to wrong direction
Route - waypoints on invalid
state
Goto Fixed - no opto found
OP 3D
Solution
If fail (open) => Replace linear
sensor cable
2. Replace opto fork
3. Replace opto comb if any of teeth
are broken.
4. Replace i3000.
See E09018
See E09018
See E09018
Software error. Take logs and boot the
device.
device.
See E27203
See E27204
See E27205
position sensor.
If fail => Replace sensor and
reposition it.
See E09018
1. Check connections:
* J3011, J1015
If fail => Replace linear and pivot
high torq cable
* J3005, J1018
If fail => Replace linear sensor cable.
Software error. Take logs and boot the
device.
1. Check connections:
* J2001, J1027