Error code
E57206
E57207
E57208
E57402
E57403
E57404
ORTHOPANTOMOGRAPH
Problem
Mixed messages
Eof
Order
Cannot re enable receiving
Extra ack
Invalid ack
™
OP 3D
Solution
4. Replace i4000
Error source i4000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
2. Replace i4000
3. Replace i3000.
Software error. Take logs and boot the
device.
Error source i3000:
1. Check connections:
* J3009, J2010
If fail => replace the main cable.
2. Replace i2000
3. Replace i3000
Error source i2000 or i3000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
* J3009, J2010
If fail => replace the main cable.
2. Replace i3000
3. Replace i2000
4. Replace i4000
Error source i4000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
2. Replace i4000
3. Replace i3000.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Error source i3000:
1. Check connections:
7 Troubleshooting
93