4 Device description
Service menu
NOTICE! The Service menu is intended only for authorized service men for
service use.
To access the Service menu from the Settings, input PIN code 1917 to the field:
NOTICE! PIN code 0612 is intended for installation only and contents of the
Service menu is limited. See the table below.
NOTICE! As a user of the OP 3D Service manual, you have a full access to the
Service menu.
In the Service menu you can:
PIN 1917
• Access the device logs and errors
(Logs)
• Update the device sofware
(Service)
• Drive the device to packing position
(Service)
• Perform motion tests(Service)
• Perform device calibrations
(Calibrations)
• View and modify configuration
preferences, for example: Define
calibration intervals, enter license
keys, select country (Preferences)
• Perform PAN and 3D tests
(Verification).
28
PIN 0612
• Drive the device to packing position
(Service)
• Perform device calibrations (Calibrations)
• View and modify configuration preferences,
for example: Define calibration intervals,
enter license keys, select country
(Preferences)
• View and modify patient size configurations
(Preferences)
• Demo mode (Preferences)
• Perform PAN and 3D tests (Verification)