4 Device description
Refer to User and Installation manual, chapter Device configuration for instructions on
setting and using the Preferences menu items.
In the Preferences > Patient Size Configuration you can:
• adjust the medium size patient values
Verification
The device offers Panoramic and 3D verification test programs with the device
movements disabled. These programs can be used for radiation related measurements.
A separate radiation measuring device/equipment is required for taking the radiation
measurements.
1. Go to Settings menudevice settings.
2. Select Service menu and enter PIN code 1917.
3. Select Verification menu that appears under the Service menu.
4.