KaVo ORTHOPANTOMOGRAPH OP 3D Service Manual page 66

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

7 Troubleshooting
Error code
E09008
E09009
E09010
E09011
E09012
E09013
66
Problem
UI panel missing
Z motor opto missing
Back panel board missing
Pivot opto missing
Linear opto missing
Rotator opto missing
Solution
1. Check connections:
* J4007, J6001
If fail (open) => replace user
interface cable
2. Replace i6000
3. Replace i4000.
1. Check connections:
* J4002, J1011, J1039
If fail (open) => replace emergency
stop with cable
* J1011, J1039
If fail (open) => replace column cable
extension
2. Replace opto fork
3. Replace i4000.
1. Check connections:
* J4013, J4101
If fail (open) => replace back panel
cable
2. Replace i4100
3. Replace i4000.
1. Check connections:
* J3005, J1026
If fail (open) => replace pivot sensor
movement cable
2. Replace opto fork
3. Replace i3000.
1. Check connections:
* J3004, J1025
If fail (open) => replace linear sensor
cable
2. Replace opto fork
3. Replace i3000.
1. Check connections:
* J2001, J1027
If fail (open) => replace linear sensor
cable
2. Replace opto fork
3. Replace i2000.
ORTHOPANTOMOGRAPH
OP 3D

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents