7 Troubleshooting
Error code
E54203
E54204
E54205
E54206
E54402
E54403
E55002
E55003
E55004
E56002
E56003
E56004
E57002
90
Problem
Event timeout
Button up during exposure
Run cannot lock settings
Z calibration without homing
Button up during start
Unhandled event
Workspace locked
Invalid state
Locked
Invalid channel
Timer init
Timer start
Failed to send message
Solution
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Error source i3000:
1. Check connections:
* J3009, J2010
If fail => replace main cable
2. Replace i2000
3. Replace i3000
Error source, i2000 or i3000:
1. Check connections:
* J4009, J3010
If fail => replace twister cable
assembly
* J3009, J2010
If fail => replace main cable
2. Replace i3000