KaVo ORTHOPANTOMOGRAPH OP 3D Service Manual page 151

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

3. Disconnect the pivot movement sensor.
4. Remove the detector of the pivot movement sensor.
5. Remove the pivot movement sensor.
6. Clean the place of the linear movement sensor with a suitable detergent.
7. Attach the new linear movement sensor in place with glue.
8. Connect the linear movement sensor.
9. Put the new pivot movement sensor in place.
10. Attach the detector of the pivot movement sensor.
11. Re-run the motion calibrations. Refer to chapters Device movement calibration and
Z-movement calibration in User and installation manual.
ORTHOPANTOMOGRAPH
NOTICE! No glue residue or grease must be left on the surface where the
linear movement sensor is attached.
NOTICE! Make sure that the sensor has been evenly attached to the suface.
OP 3D
8 Part replacement instructions
151