KaVo ORTHOPANTOMOGRAPH OP 3D Service Manual page 94

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

7 Troubleshooting
Error code
E57405
E58002
E58003
E58004
E58005
E59002
E59003
E59004
E59005
94
Problem
Failed to send message retry
Cannot get workspace
Cannot write flash
Rebooting while receiving
Timeout on receive
Start without prepared
Set when not idle
Trajectory function jammed.
Queue empty.
Trajectory function internal error Software error. Take logs and boot the
Solution
* J3009, J2010
If fail => replace the main cable.
2. Replace i2000
3. Replace i3000
Error source i2000 or i3000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
* J3009, J2010
If fail => replace the main cable.
2. Replace i3000
3. Replace i2000
4. Replace i4000
Error source i4000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable.
2. Replace i4000
3. Replace i3000.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
device.
ORTHOPANTOMOGRAPH
OP 3D

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents