4 Device description
6. Enter PIN code 1917 to access the Service menu.
7. Browse to Service> FW Update.
8. Click Select File and browse to the location to the firmware file to be loaded, for
example C:\temp\OP_3D_EMBEDDED_SOFTWARE_Rx_y_z_abcd.tar.gz. Select the
update file.
9. Click Start Update.
10. Wait for the update to finish.
11. Close the GUI to disconnect the device from the imaging software.
12. Re-connect the device by using the imaging software to finalize the update.
13. Perform all the device calibrations in accordance to the calibrations procedure, refer
to Complete device calibration chapter of the OP 3D User and Installation manual.
18
ORTHOPANTOMOGRAPH