Error code
E09014
E09015
E09016
E09017
E09018
ORTHOPANTOMOGRAPH
Problem
Coll up opto missing
Coll right opto missing
Coll left opto missing
Resistive sensor pivot missing
Resistive sensor linear missing
OP 3D
Solution
1. Check connections:
* J2005, J1034
If fail (open) => replace collimator
motors optos cable
* J1034, J1031
If fail (open) => replace collimator
2. Replace i2000.
1. Check connections:
* J2005, J1036
If fail (open) => replace collimator
motors optos cable
* J1036, J1033
If fail (open) => replace collimator
2. Replace i2000.
1. Check connections:
* J2005, J1035
If fail (open) => replace collimator
motors optos cable
* J1035, J1032
If fail (open) => replace collimator
2. Replace i2000.
1. Check that the wiper is pressing
mechanically onto the surface of an
resistive position sensor.
If fail => replace wiper, spring and
body
2. Check connections:
* J3005, J1018
If fail (open) => replace pivot sensor
movement cable
3. Replace the resistive position
sensor
4. Replace i3000.
1. Check that the wiper is pressing
mechanically onto the surface of an
resistive position sensor.
If fail => replace linear wiper
assembly
2. Check connections:
* J3004, J1017
7 Troubleshooting
67