7 Troubleshooting
Error code
E09026
E10002
E10003
E10004
E10005
E10006
E10007
E10202
70
Problem
Sensor current too high
Invalid state shift
Preheat feedback out of range
Invalid boost time
Invalid kV value
Invalid MA value
Invalid ma kV value on prepare
mA feedback out of range
Solution
If the fuse blows again => replace
i3000.
2.2 Replace i3000
1. Remove power supply connection
J2013 from i2000
1.1 Check H9 and H10 LED on i2000
board during preparation (should be
green)
If fail then
=> Replace i2000
2. Check connection: J2013, J0101
If fail then
=> Replace PAN- 3D power cable
3. Replace sensor
1.i2000:
1.1 Check fuse F2. If fuse is blown
then replace the fuse.
If the fuse blows again => replace
i2000
1.2 Replace i2000
1. Re-run THA calibrations
2. Check connections: J4019, J1005,
J7000 (pan-THA)
3. Re-run THA calibrations
4. Replace i4000
5. Re-run THA calibrations
6. Replace pan-THA
7. Re-run THA calibrations
8. Replace main cable and twister
cable assembly
9. Re-run THA calibrations.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
1. Re-run THA calibrations