8 Part replacement instructions
Removing the laser layer cover assembly
NOTICE! Remove the patient handle.
1. Remove the screw below the laser layer cover.
2. Press the back off the laser layer cover to release it.
3. Remove the cover.
Removing the lower shelf upper and lower cover
NOTICE! You need to remove the patient handle and the laser layer cover first
only if you are going to remove the lower cover. You do not need to remove the
handle before you remove the upper cover only.
NOTICE! Remove the lower shelf tray if you need to remove both the upper and
lower covers.
Required tools: Allen key 3mm
1. Remove the four screws below the lower shelf.
2. Remove the screw at the end of the lower cover.
3. Remove the upper cover half.