Error code
E28205
E28206
E28207
E28402
E28403
E31202
E31203
E31204
E31205
E31206
E32202
ORTHOPANTOMOGRAPH
Problem
Calibration missing edges
Calibration extra edges
Calibration invalid edge
Position guard failed
Timeout
Unexpected event
Invalid target
Unexpected opto hit
Edge to home opto not found
Motor driving error
Unexpected event
OP 3D
Solution
* E09013 Opto missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
1. Motion blocked externally.
2. See additional errors:
* E09013 Opto missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
1. Actuator pushed externally.
2. See additional errors:
* E09013 Opto missing
* E09019 Resistive sensor missing.
Software error. Take logs and boot the
device.
Unintended rotation of the rotating
unit.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Replace horizontal 3D laser assembly.
See E09006.
See E31204
See E09020
See E09021
See E09022.
See E09020
See E09021
See E09022.
Software error. Take logs and boot the
device.