KaVo ORTHOPANTOMOGRAPH OP 3D Service Manual page 124

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

8 Part replacement instructions
5. Detach the clamp and the cable set.
6. Turn the cable set counterclockwise to release it.
7. Pull the cable set from its locking.
8. Move the upper shelf plate slightly to make room for the cable set.
9. Slide the cable set through the hole in the upper shelf.
10. Remove the cable set.
11. Install the new cable set in its place in the carriage.
12. Lock the cable set to the position by turning the locking clamp clockwise.
13. Connect the cables to the i3000 main control board.
14. Insert and tighten the screw that holds the main cable.
15. Connect the Ethernet cable.
16. Install the i4000 board with plate and fan back.
124