Error code
E13004
E13005
ORTHOPANTOMOGRAPH
Problem
Bus-off
Stuff error
OP 3D
Solution
If fail => replace twister cable
assembly
* J3009, J2010
If fail => replace main cable
2. Replace i3000
3. Replace i2000
4. Replace i4000
Error source, i4000:
1. Check connections:
* J4009, J3010
If fail => replace twister cable
assembly
2. Replace i4000
3. Replace i3000.
Error source i3000:
1. Check connections:
* J3009, J2010
If fail => replace main cable
2. Replace i2000
3. Replace i3000
Error source, i2000 or i3000:
1. Check connections:
* J4009, J3010
If fail => replace twister cable
assembly
* J3009, J2010
If fail => replace main cable
2. Replace i3000
3. Replace i2000
4. Replace i4000
Error source, i4000:
1. Check connections:
* J4009, J3010
If fail => replace twister cable
assembly
2. Replace i4000
3. Replace i3000.
Error source i3000:
1. Check connections:
7 Troubleshooting
75