KaVo ORTHOPANTOMOGRAPH OP 3D Service Manual page 91

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

Error code
E57003
E57004
E57005
E57006
E57007
E57202
ORTHOPANTOMOGRAPH
Problem
Received message with no
service registered
Isr invalid service
Isr frame queue full
Isr interupt enabled failed
Buffer full force reboot
ACK timeout
OP 3D
Solution
3. Replace i2000
4. Replace i4000
Error source, i4000:
1. Check connections:
* J4009, J3010
If fail => replace twister cable
2. Replace i4000
3. Replace i3000.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Error source i3000:
1. Check connections:
* J3009, J2010
If fail => replace the main cable
2. Replace i2000
3. Replace i3000
Error source i2000 or i3000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable
* J3009, J2010
If fail => replace the main cable
2. Replace i3000
3. Replace i2000
4. Replace i4000
Error source i4000:
1. Check connections:
* J4009, J3010
If fail => replace the twister cable
2. Replace i4000
7 Troubleshooting
91

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents