8 Part replacement instructions
2. Turn the adjustment screw to adjust the height of the sensor. One turn is equivalent
of 0.7 mm.
3. Tighten the two vertical screws when you have done the adjustment.

8.3.3 Replacing the horizontal lasers

NOTICE! You need to remove the carriage right and left cover before this
procedure.
NOTICE! Turn the upper shelf aside to a position that allows you to access the
carriage easily.
Required tools: Allen key 3mm
1. Remove the three screws that hold the protective plate.
2. Disconnect the mirror opto cable.