Error code
E10210
E10211
E10402
E10403
E11002
E11003
E11004
E11005
E11006
E11202
E11203
E11204
E11205
E11206
ORTHOPANTOMOGRAPH
Problem
Calibration feedback flat on scan 1. Check connections: J4005, J4019,
Program unexpected event
Tube fail recovering
Calibration Unexpected event
Timeout on ack
Invalid panel mode
Unknown event
Unknown reply
Receive buffer full
Too short reply
CRC check failed
Invalid token count
Invalid cropping settings
Capturix Failed
OP 3D
Solution
J1004, J1005, J7000 and J7002 (pan
THA)
2. Re-run THA calibrations
3. Replace i4000
4. Re-run THA calibrations
5. Replace pan THA
6. Re-run THA calibrations
7. Replace main cable and twister
cable assembly
8. Re-run THA calibrations.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
1. Re-program i2000 board
2. Replace i2000 board.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.