KaVo ORTHOPANTOMOGRAPH OP 3D Service Manual page 108

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

8 Part replacement instructions
2. Remove the screws (2 pcs) from below of the rotating unit.
3. Pull the rotator front and back covers off.
Removing the tube head back cover
Required tools: Allen key 3mm
NOTICE! The tube head back cover is attached with two screws from top of the
rotating unit and with one screw on the tube head inner cover.
1. Remove or loosen the screws from right and left side of the rotating unit.
2. Remove the screw on the tube head front cover.
3. Remove the tube head back cover.
Removing the tube head front cover
Required tools: Allen key 3mm
108
ORTHOPANTOMOGRAPH
OP 3D

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents