KaVo ORTHOPANTOMOGRAPH OP 3D Service Manual page 69

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

Error code
E09023
E09024
E09025
ORTHOPANTOMOGRAPH
Problem
Mirror opto missing
24V power supply voltage out of
range
24V motor voltage out of range
OP 3D
Solution
* J2002, J1012
If fail (open) => replace rotating high
torq mot cable
2.2 Replace motor
2.3 Replace i2000
3 For sensor swap:
3.1 Check connections:
* J2003, J1014
If fail (open) => Replace detector
swap cable
3.2 Replace motor
3.3 Replace i2000
4 For collimator:
4.1 Check connections:
* J2005, J1034, J1035, J1036
If fail (open) => replace collimator
motors optos cable
4.2 Replace collimator
4.3 Replace i2000
5 For laser lift motion:
5.1 Replace horizontal 3D laser
assembly
5.2 Replace i4000.
1. Check connections:
* J4025, J1023
If fail (open) => replace mirror opto
cable
2. Replace opto fork
3. Replace i4000.
Replace i4000.
1.i2000:
1.1 Check fuse F2
If the fuse is blown replace the fuse.
If the fuse blows again => replace
i2000.
1.2 Replace i2000
2. i3000:
2.1. Check fuse F1
If the fuse is blown replace the fuse.
7 Troubleshooting
69

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents