Error code
E64002
E64003
E64202
E64203
E90002
E90003
E90004
E90005
E90006
E91202
E92202
E99602
E99603
E99604
E99605
E99606
98
Problem
Init failed
Start failed
Control loop too slow
Watchdog fail
No error
Invalid command
Invalid arguments
Resource not available
Error while executing
Network speed too low
Phy register write error
Debug point 0
Debug point 1
Debug point 2
Debug point 3
Debug point 4
Solution
if fail => Replace i4000.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
1. Check the Ethernet link partner
(MWS) switch or PC settings support
Gigabit Ethernet.
2. Check that the Ethernet cabling is
CAT5e or CAT6.
3. Replace i3000.
1. Restart the device.
If fail => replace i2000.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.