8 Part replacement instructions
4. Remove the screw from the cable clamp and turn the clamp up.
5. Move the rotating unit so that it is in a 90 degree angle with the upper shelf and
push it against the column.
NOTICE! The tube head must be in home position (on the right hand side of
the device).
6. Pull to remove the cable clamps on both ends of the rotating unit.
126