Cleaning Or Replacing The Fresh Water Tank - KaVo STERIclave B 2200 Operating Instruction

Table of Contents

Advertisement

Operating Instruction STERIclave B 2200 / 2200 P
|
7 Maintenance

7.5 Cleaning or replacing the fresh water tank

+
7.5 Cleaning or replacing the fresh water tank
Requirement
The unit must be switched off. "OFF" appears on the display.
▶ Empty the fresh water tank. To do this, connect the drain hose to the right-hand
quick connection and drain the water into a container (min. capacity 2.6 litres).
▶ Use the filter key to remove the fresh water filter.
▶ Use compressed air or ultrasound to clean the filter.
or
▶ Change the filter.
▶ Carefully screw the filter back into place by hand.
▶ Fill the tank with high-quality distilled or demineralized water.
▶ Press 2 and the Power button at the same time.
The water level will be topped up automatically.
The Ready LED will light up as soon as the water reaches a sufficient level. The
unit is now ready for operation.
63/78