7 Troubleshooting
Error code
E18202
E18204
E18205
E18206
E18402
E18404
E18405
E19002
E19003
E19004
80
Problem
Wrong state
Drive aborted timeout
Opto not found
No pulses
Overheat
Proximity disabled or no pulses
Drive aborted button
Self-calibration failed
Invalid state shift
Self-diagnostic failed
Solution
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
1. Check connections:
* J4002, J1011, J1039
If fail (open) => replace emergency
stop with cable
* J1011, J1039
If fail (open) => replace cable column
cable extension
2. Replace opto fork
3. Replace i4000.
1. Check connections:
* J1037, J4002
If fail => Replace column cable
extension
2. Replace i4000
3. Replace column.
Let the device cool down.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
1. Check that there are no objects on
the UI panel.
If fail => remove object
2. Replace the upper cover of the
lower shelf (contains the UI panel
module).
Software error. Take logs and boot the
device.
1. Check connections:
* J4007, J6001
If fail => replace user interface cable
2. Replace the upper cover of the
lower shelf (contains the UI panel
module)
3. Replace i4000.
ORTHOPANTOMOGRAPH
OP 3D