Service
In the Service menu you can:
• update the device firmware
• perform accelerated motion test for different main movents
• test the door and mirror door switch status change
• drive the device to packing position
• test the 3D sensor connectivity
• test the functionality of internal communication buses
• check the statuses of the internal power supplies.
For more information on how to update the device firmware, see chapter
update.
Motion tests are used for testing the individual movements of the device. To perform a
motion test to the following:
• Select the main movement (PIVOT, ROTATION, LINEAR) or other movement (LASER,
UPPER COLLIMATOR, LEFT COLLIMATOR, RIGHT COLLIMATOR)
• For main movements, select the number of movements for the test.
• For other movements, select the range of movement (min/max). Change the range
between tests when you test the same movement.
• Press the exposure button to perform the movement.
• Verify the selected movement visually.
Internal communication bus test results and power supply status results are provided in
the system logs. After you have executed the test, check the results from Logs menu by
clicking log/messages.
Logs
In the Logs menu you can access device logs and error messages.
To dowload logs or error messages to the workstation, select the logs/error messages
and click Download.
ORTHOPANTOMOGRAPH
OP 3D
4 Device description
Firmware
29