7 Troubleshooting
Error code
E10206
E10207
E10208
E10209
72
Problem
Calibration feedback constant
Calibration feedback not raising
enough
Calibration invalid pwm
Calibration feedback invalid on
scan
Solution
6. Re-run THA calibrations
7. Replace main cable and twister
cable assembly
8. Re-run THA calibrations.
1. Check connections: J4005, J4019,
J1004, J1005, J7000 and J7002 (pan
THA)
2. Re-run THA calibrations
3. Replace i4000
4. Re-run THA calibrations
5. Replace pan THA
6. Re-run THA calibrations
7. Replace main cable and twister
cable assembly
8. Re-run THA calibrations.
1. Check connections: J4005, J4019,
J1004, J1005, J7000 and J 7002 (pan
THA)
2. Re-run THA calibrations
3. Replace i4000
4. Re-run THA calibrations
5. Replace pan THA
6. Re-run THA calibrations
7. Replace main cable and twister
cable assembly
8. Re-run THA calibrations.
Software error. Take logs and boot the
device.
1. Check connections: J4005, J4019,
J1004, J1005, J7000 and J7002 (pan
THA)
2. Re-run THA calibrations
3. Replace i4000
4. Re-run THA calibrations
5. Replace pan THA
6. Re-run THA calibrations
7. Replace main cable and twister
cable assembly
8. Re-run THA calibrations.
ORTHOPANTOMOGRAPH
OP 3D