KaVo ORTHOPANTOMOGRAPH OP 3D Service Manual page 79

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

Error code
E13011
E13012
E13013
E14002
E14003
E14402
E14403
E14404
E15002
E15003
ORTHOPANTOMOGRAPH
Problem
FIFO0 overrun
FIFO1 overrun
Unkown status
Handle locked
Diag listener failed
HAL error callback
Operation read failed
Operation write failed
Flash read failed
Flash data verify failed
OP 3D
Solution
3. Replace i3000
Error source, i2000 or i3000:
1. Check connections:
* J4009, J3010
If fail => replace twister cable
assembly
* J3009, J2010
If fail => replace main cable
2. Replace i3000
3. Replace i2000
4. Replace i4000
Error source, i4000:
1. Check connections:
* J4009, J3010
If fail => replace twister cable
assembly
2. Replace i4000
3. Replace i3000.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
Software error. Take logs and boot the
device.
7 Troubleshooting
79

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents