KaVo ORTHOPANTOMOGRAPH OP 3D Service Manual page 85

Hide thumbs Also See for ORTHOPANTOMOGRAPH OP 3D:
Table of Contents

Advertisement

Error code
E26203
E26204
E26205
E26206
E26207
E26402
E26403
E27202
E27203
ORTHOPANTOMOGRAPH
Problem
Unexpected opto hit
Missed opto
Calibration missing edges
Calibration extra edges
Calibration invalid edge
Position guard failed
Timeout
Position fail
Unexpected opto hit
OP 3D
Solution
1. Actuator pushed externally.
2. See additional errors:
* E09011 Opto missing
* E09017 Resistive sensor missing.
1. Motion blocked externally.
2. See additional errors:
* E09011 Opto missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
1. Motion blocked externally.
2. See additional errors:
* E09011 Opto missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
1. Actuator pushed externally.
2. See additional errors:
* E09011 Opto missing
* E09017 Resistive sensor missing.
Software error. Take logs and boot the
device.
Unintended angle change of the upper
shelf.
Software error. Take logs and boot the
device.
1. Pushed/blocked externally.
2. See additional errors:
* E09012 Opto missing
* E09018 Resistive sensor missing
* E09020 Stepper thermal warning
* E09021 Stepper thermal shutdown
* E09022 Stepper open coil.
1. Actuator pushed externally.
2. See additional errors:
* E09012 Opto missing