Commissioning
6.5 Quick Start Wizards
6.5.1

Setting wizards via graphical display

1. Press RIGHT arrow
2. Choose Wizards (1.) (Page 59) > Quick Start (1.1) (Page 61), and then the appropriate
quick start:
– Quick Start Level (1.1.1.) (Page 61)
– Quick Start Volume (1.1.2.) (Page 66)
– Quick Start Flow (1.1.3.) (Page 73)
– Pump Control (1.2.) (Page 83)
3. At each step, press DOWN arrow
next item, or RIGHT arrow
4. In Edit mode, scroll to desired item and press RIGHT arrow
press DOWN arrow
5. Repeat steps 3 and 4 until you complete all the settings and get a prompt to configure
another measurement point or relay (for Pump Control wizard). Pressing
you set another point or relay. Pressing
wizard.
6. Press
then return to PROGRAM menu. Press
While configuring the device through the wizards, you can press UP arrow
back, or LEFT arrow
Notes:
● The Quick Start Wizard settings are inter-related and changes apply only after you
choose Finish in the final step.
● Perform customization for your application only after the Quick Start has been completed.
● The following are key terms used throughout the QSW and Parameters:
Default: the factory-set value or option; indicated with an asterisk (*) or specified as a
preset value.
Global: pertains to values that are common for all inputs and outputs on the unit.
Index: when parameters apply to more than one input, they are indexed. The index
selector value defines the input/output for that parameter. For example, index relates to
transducer inputs or mA outputs, and can also refer to relays, communications ports, and
other parameters.
60
to enter PROGRAM mode.
to open Edit mode: the current selection is highlighted.
to continue.
to FINISH and apply the settings made in that particular wizard. The display will
to cancel.
to accept default values and move directly to the
for NO takes you to the end of the chosen
to return to Measurement mode.
Operating Instructions, 06/2018, A5E36281317-AC
to store the change, then
for YES lets
to go one step
HydroRanger 200 HMI