To change the relay assignment, see Relay selector (3.2.9.1.) (Page 260).
8.7.2.10

Temperature peak values (3.2.10)

The following features display the high and/or low temperatures in °C.
If the unit is powered up without a temperature sensor connected, the value – 50°C is
displayed. This information can help trace problems with both built-in and external
temperature sensors.
Transducer selector (3.2.10.1.)
Sets the transducer index for all parameters applicable to this sub-menu.
Transducer temperature maximum (3.2.10.2.)
Shows the highest temperature encountered, as measured by the temperature sensor in the
transducer (if applicable).
Index
Values
Related
To reset the log after a short circuit on the transducer wiring, enter the value -50.
Transducer temperature minimum (3.2.10.3.)
Shows the lowest temperature encountered, as measured by the temperature sensor in the
transducer (if applicable).
Index
Values
Related
To reset the log after an open circuit on the transducer wiring, enter the value 150.
TS-3 temperature maximum (3.2.10.4.)
Shows the highest temperature encountered, as measured by the TS-3 temperature sensor
(if applicable).
Index
Values
Related
To reset the log after a short circuit on the transducer wiring, enter the value -50.
HydroRanger 200 HMI
Operating Instructions, 06/2018, A5E36281317-AC
Transducer
Range: - 50 ... 150 °C (Read only)
Preset: - 50 °C
Transducer Temperature Minimum (3.2.10.3.) (Page 261)
Transducer
Range: - 50 ... 150 °C (Read only)
Preset: 150 °C
Transducer temperature maximum (3.2.10.2.) (Page 261)
Global
Range: - 50 ... 150 °C (Read only)
Preset: - 50 °C