Power-On Resets (3.2.4.); Power-On Time (3.2.5.); Relay Selector (3.2.6.); Relay Logic Test (3.2.7.) - Siemens HydroRanger 200 HMI Operating Instructions Manual

Sitrans ultrasonic controllers
Hide thumbs Also See for HydroRanger 200 HMI:
Table of Contents

Advertisement

8.7.2.4

Power-ON resets (3.2.4.)

Displays the number of times power has been applied since Manufacture Date (3.1.9.)
(Page 253).
Index
Values
Related
8.7.2.5

Power-ON time (3.2.5.)

Displays the number of days this device has been in operation.
Index
Values
Related
The RUN Time value is updated once a day, and cannot be reset. However, in the event of a
power interruption, the counter won't advance. Therefore, a unit that is powered down on a
regular basis will not have an accurate value.
8.7.2.6

Relay selector (3.2.6.)

Sets the relay index for all parameters applicable to this sub-menu.
8.7.2.7

Relay logic test (3.2.7.)

Forces the relay control logic into an activated or de-activated state.
Index
Options
Related
This parameter tests site wiring and control logic programming. Forcing the relay to an
activated or de-activated state is similar to the device detecting an event and responding to
it. Helpful in testing new installations and diagnosing control problems.
8.7.2.8

Measurement values (3.2.8.)

Level selector (3.2.8.1.)
Sets the Level Point index for all parameters applicable to this sub-menu.
HydroRanger 200 HMI
Operating Instructions, 06/2018, A5E36281317-AC
Global
Range: 1 ... 9999 (Read only)
Manufacture date (3.1.9.) (Page 253)
Power-ON time (3.2.5.) (Page 255)
Global
Range: 1 ... 9999 (Read only)
Manufacture date (3.1.9.) (Page 253)
Power-ON resets (3.2.4.) (Page 255)
Relay
*OFF
Activate relay control
De-activate relay control
Relay function (2.8.1.4.) (Page 195)
Parameter reference
8.7 Maintenance and Diagnostics (3.)
255