Commissioning
6.5 Quick Start Wizards
Response rate
Sets the reaction speed of the device to measurement changes in the target range.
Note
• Any changes made to Fill rate/minute (2.3.2.) (Page 173), Empty rate/minute (2.3.3.)
(Page 173), or Response rate (2.3.4.) (Page 174) parameters following the completion of
the wizard will supersede the Response Rate setting.
• Response Rate always displays in meters per minute (m/min).
Index
Options
Factory setting
Use a setting just faster than the maximum filling or emptying rate (whichever is greater).
Slower settings provide higher accuracy, faster settings allow for more rapid level
fluctuations.
Dimension A
The height of the vessel bottom when the bottom is conical, pyramidal, parabolic, spherical,
or flat-sloped.
Dimension A as used in Vessel shape (2.7.2.) (Page 186).
Index
Values
Enter ONE of the following:
● Height of the tank bottom, if Vessel Shape (2.7.2.) (Page 186) = Conical Bottom,
Parabolic Bottom, Half Sphere Bottom, or Flat Sloped Bottom
● Length of one end section of the tank, if Vessel Shape (2.7.2.) (Page 186) = Parabolic
Ends, in Units (2.1.1.) (Page 162)
70
Transducer
Slow (0.1 m/min)
*Medium (1.0 m/min)
Fast (10.0 m/min)
Medium (1.0 m/min)
Level
Range: 0.0 ... 99.00 m or equivalent, depending on Units (2.1.1.) (Page 162).
Preset: 0.000
HydroRanger 200 HMI
Operating Instructions, 06/2018, A5E36281317-AC