3.4.2.3
Patient Applied Risk Current

Table 3-3: Enclosure Leakage Current Limits

AC LINE
NEUTRAL
CORD
LINE
CORD
Closed
Closed
Closed
Closed
Closed
Open
Open
Closed
Open
Open
Open
Closed
This test is in compliance with AAMI Standard ES1 (patient applied risk
current), and IEC 601-1 (patient auxiliary current). The leakage current is
measured between any individual patient connection and power (earth) ground.
The applied voltage for AAMI/ANSI is 120 VAC 60 Hz, and for IEC 601-1 the
applied voltage is 264 VAC 50 to 60 Hz.
1.
Configure the electrical safety analyzer as follows:
Function:
Range: µA
2.
Connect the monitor AC plug to the electrical safety analyzer as
recommended by the analyzer operating instructions for Patient Leakage
Current.
3.
Connect the patient leakage input lead of the electrical safety analyzer to all
pins of the monitor's patient cable at the end of the cable.
4.
The equipotential terminal is not connected to ground.
5.
All functional earth terminals are not connected to ground.
6.
Measure the leakage current between the patient connector and earth. See
Table 3-4.

Table 3-4: Patient Leakage Current Limits

AC LINE
NEUTRAL
POLARITY
LINE
Normal
Closed
Normal
Open
Normal
Closed
Reverse
Closed
Reverse
Open
Reverse
Closed
Section 3: Performance Verification
POWER
LINE
GROUND
CABLE
Closed
Open
Closed
Closed
Closed
Open
Patient Leakage
POWER
LINE
GROUND
CABLE
Closed
Closed
Open
Closed
Closed
Open
IEC 601-1
AAMI/ANSI
ES1
STANDARD
100 µA
100 µA
500 µA
300 µA
500 µA
300 µA
500 µA
100 µA
500 µA
300 µA
500 µA
300 µA
IEC 601-1
AAMI/ANSI
ES1
STANDARD
100 µA
10 µA
500 µA
50 µA
500 µA
50 µA
100 µA
10 µA
500 µA
50 µA
500 µA
50 µA
3-11