Transport Regulator/Cap Assembly - Mallinckrodt INOmax DSIR Pocket Manual

Delivery system, series 3 software
Hide thumbs Also See for INOmax DSIR:
Table of Contents

Advertisement

Transport Regulator/
Cap Assembly

Transport Regulator/Cap Assembly

WARNING:
• A new INOMAX cylinder and regulator must be purged
before use to ensure the patient does not receive an excess
level of NO
• Loss of communication between the INOmax DS
INOMAX cylinder for more than one hour will result in
interruption of INOMAX delivery.
Caution: When using the Transport Regulator/Cap Assembly
(PN 10022) ensure the cap is in place on the cylinder
and the infrared cable is connected to the infrared
connector port on the back of the INOmax DS
Note: Check the INOMAX cylinder for the correct product identity
labels, cylinder concentration and expiration date. Ensure
the INOMAX gas cylinder has more than 500 psig.
Step One
Note: Ensure the white plastic tip is in place on the regulator
connector and not chipped or cracked (see Figure 2).
Connect a high
pressure regulator
to an INOMAX
cylinder and tighten
the fitting to the
INOMAX cylinder
(see Figure 1).
Part No. 20751 Rev-01
2014-08
.
2
Figure 1
Figure 2