Vapotherm Precision Flow - Mallinckrodt INOmax DSIR Pocket Manual

Delivery system
Hide thumbs Also See for INOmax DSIR:
Table of Contents

Advertisement

Connection to the Vapotherm Precision Flow
• The INOmax DS
proportion to the NO setting (up to 10% at 80 ppm) and subtracts gas
from the breathing circuit via gas sampling at a nominal flow rate of
0.23 L/min.
• These effects impact the delivered gas flow rate when using the
Vapotherm Precision Flow. It is recommended that after an NO setting
change the user checks the delivered gas flow rate and adjusts the
gas source flow rate as necessary.
• Follow all manufacturer instructions for connection to the Vapotherm
Precision Flow.
1
2
8
9
7
adds NO/N
gas flow to the breathing circuit flow in
IR
2
3
6
5
Circuit Connection
4
1. Patient Gas Sample Line
with Nafion
2. INOmax DSIR
3. Precision Flow Unit
4. Injector Module
5. Patient Delivery Tube
6. Oxygen Tubing Sample Tee
7. Patient Cannula
8. Injector Module Electrical Cable
9. NO/N2 Injector Tube
Part No. 20728 Rev-01
47
Diagrams
2012-09

Hide quick links:

Advertisement

Table of Contents