Analog Output Configuration
(Select Action)

Select Range

Thermo Fisher Scientific
mA, or 4 mA. By default, this parameter is set to on and 5% over and
under range is allowed for all analog output channels.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > Allow Over/Under Range.
ALLOW OVER/UNDER RANGE:
CURRENTLY:
SET TO:
RANGE
AVG
The Analog Output Configuration menu displays a list of the analog
output configuration choices, from which the user selects the parameter to
adjust for the selected output channel. Configuration choices include
selecting range, setting minimum/maximum values, and choosing the
signal to output.
Note The settings are different depending on the AUTO/MANUAL
MODE and RANGE MODE settings. Changing either of these settings
will produce a different set of analog output data items. If either of these
modes is going to be changed on a regular basis, then the user must
configure each set of data separately after changing the modes to each
setting.
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > All Voltage Channels, All Current
Channels, Voltage Channel 1–6 or Current Channel 1–6.
ANALOG OUTPUT CONFIG:
>SELECT RANGE
SET MINIMUM VALUE
SET MAXIMUM VALUE

CHOOSE SIGNAL TO OUTPUT

RANGE
AVG
The Select Range screen is used to select the hardware range for the selected
analog output channel. Possible ranges for the voltage outputs are: 0-100
mV, 0-1, 0-5, and 0-10 V. Possible ranges for the current outputs are: 0-20
mA and 4-20 mA.
ON
OFF
TOGGLE VALUE
DIAGS ALARM
DIAGS ALARM
Operation
Instrument Controls Menu
?
Model 42i Instruction Manual
3-49