Collect Data; Post Wash - Thermo Scientific Orion 2230XP User Manual

Silica analyzer
Table of Contents

Advertisement

2230XP Silica Analyzer
To edit the Reagent volume, from the MAIN MENU follow the sequence:
MAIN MENU ➔ SETUP ➔ METHOD ➔ (+) to go to page 2 ➔STIR TIME (1,
2, 3)
During the addition of reagent, the stirrer is on and the info field (if enabled) shows the reagent
pumping progress, with a bar and the volume pumped. After adding the reagent, the countdown
of stir time is also shown along with the absorbance of the sample mixture.
If the pump is not assigned to a reagent, it will abort the measurement and the
NOTE:
info field will flash with the error message: "PUMP IS NOT ASSIGNED".

Collect Data

In collect data, the final reading is recorded and the final absorbance is displayed at the end of
the cycle.
The stability of the recorded reading depends on the specified stability criteria. You have the
choice of editing the stability criteria. To do this from the MAIN MENU follow the sequence:
MENU ➔ SETUP ➔ ANALYSIS PARAMETERS ➔STABILITY CRITERIA
Now the concentration is calculated (as follows) and the current concentration is updated.
Concentration = (Slope x Absorbance) + Intercept
(Where slope and intercept are from the CURRENT calibrated values of the instrument for
measurement.)
If unstable and doesn't meet the stability criteria, the info field will flash with the
NOTE:
warning message "COLLECTING DATA, UNSTABLE SIGNAL".

Post Wash

The programmable POST WASH is to wash the reaction cell after the measurement and
evacuate any residue of the reaction mixed solution. This event is optional and can be disabled,
or the reaction cell can be washed with selectable sources with editable volumes and stir times,
with up to five programmable steps. To perform post wash after measurement, select
ANALYSIS PARAMETERS for the POST WASH EVENT.
To set the post wash parameters from the MAIN MENU follow the sequence:
MAIN MENU ➔ SETUP ➔ ANALYSIS PARAMETERS ➔ (+) POST WASH
EVENT ➔ POST WASH CYCLES
Calibration of the Instrument for Measurement Analysis
Section 6
Thermo Scientific Orion
47

Advertisement

Table of Contents
loading

Table of Contents