Operating the Instrument

Collecting Data

18
Nicolet iS5 LS User Guide
When the workflow is completed, the software displays the report. If the instrument
passed, a green check mark appears to the right of each test result and the operator can
proceed. Here is an example.
Figure 15. Example Calibration Check report
8. If your instrument passed the performance test, click continue to complete the workflow.
If your report includes a red "X," rerun the Calibration Check workflow. If it fails again,
the resolution depends on which test has the red X. For more information, refer to the
Performance test failed
9. To print the report, click the Print button.
Note To retrieve data from a Calibration Check workflow, use Data Hub.
Related Topics
Daily Performance Test
Instrument Performance Tests
This section explains how samples are analyzed with the instrument. These topics are covered:
Sampling Tips
Background Measurement
Sample Measurements
topic in the Troubleshooting section of the instrument user guide.