Operation
Diagnostics Menu
Relay States
Test Analog Outputs
Set Analog Outputs
3-64
Model 42i Instruction Manual
DIGITAL INPUTS:
>INPUT 1
INPUT 2
INPUT 3
INPUT 4
INPUT 5
RANGE
AVG
The Relay States screen displays the state of the 10 digital outputs and
allows toggling of the state to either on (1) or off (0). The relays are
restored to their original states upon exiting this screen.
In the Main Menu, choose Diagnostics > Relay States.
●
Press
to toggle and set the relay state to open or closed for the
●
selected digital output.
RELAY STATE:
>OUTPUT 1
OUTPUT 2
OUTPUT 3
OUTPUT 4
OUTPUT 5
RANGE
AVG
The Test Analog Outputs menu is used to set each of the analog output
channels to zero or full-scale. Channel choices include all analog outputs, 6
voltage channels, and 6 current channels (if the I/O expansion board
option is installed).
In the Main Menu, choose Diagnostics > Test Analog Outputs.
●
TEST ANALOG OUTPUTS:
>ALL
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
VOLTAGE CHANNEL 4
RANGE
AVG
The Set Analog Outputs screen contains three choices: Set to full-scale, set
to zero, or reset to normal. Full-scale sets the analog outputs to the full-
scale voltage, zero sets the analog outputs to 0 volts, and reset returns the
output to normal operation. The analog outputs are returned to normal
1
1
1
1
1
DIAGS ALARM
0
0
0
1
1
DIAGS ALARM
DIAGS ALARM
Thermo Fisher Scientific