Dut Measurement; Dut Measurement Using Pin Probe - Agilent Technologies 42941A Operation And Service Manual

Impedance probe kit
Table of Contents

Advertisement

Operation

DUT Measurement

DUT Measurement
Before performing DUT measurement, open and short compensation should be
done as described in the previous sections.

DUT Measurement Using Pin Probe

When measuring implemented circuits, mounted devices, and printed circuit
patterns, use the pin probe.
Figure 3-7
DUT Measurement Using Pin Probe
NOTE
The pin of the pin probe is replaceable. Replace it when damaged or dirty.
Adjusting Pin-to-GND Gap
The gap between the pin and the GND contact is adjustable for DUT, ranging from
0.5 mm to 13.5 mm. To adjust the gap, release the screw fastening the GND
contact and then rotate the contact (Figure 3-8).
The probe is grounded 25 Ω ( nominally). Do not connect the probe directly to the
CAUTION
circuit with a DC output. If you connect a circuit with a DC output, resulting in as
follows.
In the case of measuring a circuit that could output a DC signal, connect an
24
It may cause the probe to be damaged.
It may influence the circuit to not perform correctly or may damage it entirely.
Measurement value will be incorrect.