than a lower limit value is lower than a certain defined range,
an alarm will be given.
At the same time, the alarm also has a feature to delay.
Event refers to marking the current action, including
measurement, calibration and deep cleaning.
Pre-running/post-running Functions: According to
different site applications, when other external equipment is
required to run for a period before triggering the Ammonia
Analyzer for analysis, the pre-run function must be initiated
(e.g. in the event that a sample conditioning system is
installed).
Similarly, other external device need to be triggered and
started after the analysis of the Online Ammonia Analyzer is
completed, the post-run function must be started.
Pre-running: Select "Parameter Setting" >> "Input/Output"
>> "Pre-/Post-Running" to set the pre-running command.
Post-running: The post-running setting is similar to the
pre-running setting. Select "Parameter Setting" >>
"Input/Output" >> "Pre-/Post-running" to set post-running
commands.
Digital Input: It is used to set the functions and delay.
Functions include measurement, deep cleaning, calibration,
e m p t yi n g a n d s to p or n o n e . W h e n i t i s s e t t o t h e
corresponding function, the corresponding function is started
when there is a input. The delay is used to set a specific time
delay after input and then execute the corresponding
function. This instrument provides two digital inputs, which
can be configured separately.
Analog Output: It is used for 4 - 20 mA external output. It
can be set to OFF or 4 - 20 mA. When set to 4 - 20 mA, the
corresponding values of 4 mA (lower limit) and 20 mA (upper
limit) need to be set. This instrument provides two digital
outputs which can be configured separately.
MODBUS Output (RS485): RS485 is used to set the station
address, baud rate, calibration mode, data bit and stop bit.

Record Query

Record query is used to query various records, including
alarm records, measurement records, calibration records and
operation records. Click different record bars to enter different
record query interfaces.
Orion 8010cX Ammonia Analyzer
Alarm Records
Alarm records are used to record various alarms and fault
events occurring during the operation process of the
instrument, including the time, date, event, current status of
event and alarm code. When querying alarm records, you can
query based on time (including 1 week, 1 month, 3 months, 6
months, 1 year and all). Alarm records can be exported when
you are authorized. When exporting, a designated type of
USB storage device must be connected to the instrument.
Currently-supported USB types: USB devices of fat32
file system. It is recommended to use standard USB
devices from Thermo Fisher Scientific or USB devices
(less than 64G) from SanDisk and Kingston brands.
Measurement Records
Measurement record query is used to view and export test
results, inc luding the me asure ment ti me, date, and
measurement details, etc. Click the detailed information of
each measurement to view the measurement status, analysis
time, concentration, sample absorbance, measuring range
and other specific information. When querying measurement
records, you can query based on time (including 1 week,
1 month, 3 months, 6 months, 1 year and all). When you are
authorized, you can export measurement records. When
exporting, a designated type of USB storage device needs to
be connected to the instrument.
Calibration Records
Calibration record query is used to view and export calibration
results, including the time, date, and calibration information.
Click each calibration information once to view detailed
recording parameters, including the calibration status,
calibration time, calibration concentration, blank absorbance,
absorbance of standard solution, intercept, slope and
calibration range. When querying calibration records, you can
query based on time (including 1 week, 1 month, 3 months, 6
months, 1 year and all). When you are authorized, you can
export calibration records. When exporting, a designated type
of USB storage device is required to be connected to the
instrument.