Table B-4. Record Output Formats - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

Thermo Fisher Scientific
Send:
lrec format
Receive:
lrec format 1
set erec format format
set lrec format format
set srec format format
These commands set the output format for long and short records, and
dynamic data, according to
record output format to ASCII with text.
Send:
set lrec format 1
Receive:
set lrec format 1 ok
Table B–4. Record Output Formats
Output Format
Format
0
ASCII no text
1
ASCII with text
2
Binary data
erec layout
lrec layout
srec layout
These commands report the layout (string indicating the data formats) for
data that is sent out in response to the erec, lrec, srec, and related
commands. For details on how to interpret the strings, see "Record Layout
Definition" later in this appendix.
Send:
lrec layout
Receive:
lrec layout %s %s %lx %f %f %f %f %f %f %f %f %f %f %f
%f
t D L ffffffffffff
flags no nox hino hinox pres pmtt intt rctt convt smplf
ozonf pmtv
lrec mem size
srec mem size
These commands report the number of lrecs and srecs that can be stored
with the current settings and the number of blocks reserved for lrecs and
srecs. The example that follows shows that 1075 blocks were reserved for
lrecs and the maximum number of lrecs that can be stored in memory is
241979. Memory allocation can be changed using the malloc command.
Send:
lrec mem size
Receive:
lrec mem size 241979 recs, 1075 blocks
Table
B–4. The example below sets the long
Model 42i Instruction Manual
C-Link Protocol Commands
Datalogging
B-23

Advertisement

Table of Contents
loading

Table of Contents