Preventive Maintenance
Replacing Water Trap Probe Filter and O-Rings
Replacing Water
Use the following procedure to replace the water trap probe filter
(Figure
5–4).
Trap Probe Filter
and O-Rings
1. Unscrew the bottom housing from the top housing.
2. Remove filter.
3. Place new filter into cavity of top housing.
4. Filter can be inserted in either direction into top housing.
5. Screw bottom housing into top housing.
Probe Nut
Nozzle
O-ring
Water
Trap
O-ring
Nozzle Base
Figure 5–4. Water Trap Probe Assembly – Filter and O-Ring Replacing