Servicing
Ozonator Assembly Replacement
7-22
Model 42i Instruction Manual
2. Carefully disconnect the plumbing at the glass inlet and outlet of the
ozonator.
3. Disconnect the stainless steel tubing from the flow transducer.
4. Loosen the four captive screws securing the ozonator bracket to the
floor plate.
Captive Screws (4)
Ozonator
Ozonator Transformer
Screws (4)
Figure 7–12. Replacing the Ozonator Assembly
5. Remove the two screws securing the ozonator to the ozonator bracket.
6. Unplug the ozonator from the ozonator transformer by lifting the
ozonator straight up.
7. To install the ozonator, follow the previous steps in reverse.
Ozonator Assembly – Top View
Ozonator Assembly – Side View
Flow Transducer
Ozonator Bracket Screws (2)