49iQ Photometer DMC
Detector Assembly
Replacement
7-34
49iQ Instruction Manual
Use the following procedure to remove the pre-amplifier/detector assembly.
Equipment required:
Hex drive, 7/64
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2–1).
2. Unplug electrical cable from detector 1 and detector 2.
Figure 7–40. Preamp Detector Cover Removal
3. Using a 7/64 hex drive, unfasten two #6-32 cap screws.
Thermo Scientific