CAUTION
Figure 2-2
WARNING
To Make an Interlock Circuit
The interlock circuit is designed to prevent electrical shock when a user touches the
measurement terminals.
You must install an interlock circuit on a shielding box to prevent hazardous
voltages when the door of the shielding box is open.
Figure 2-2 shows the pin assignments of the interlock connector mounted on a
connector plate or test fixture.
Interlock Connector Pin Assignments
Potentially hazardous voltages may be present at the Force, Guard, and Sense
terminals when the interlock terminals are shorted.
Agilent 4155C/4156C User's Guide Vol.1, Edition 11
Installation
Mounting Connectors
2-33