Table 8-5
Voltage
Set
Range
Reso.
± 2 V
100 μV
± 20 V
1 mV
± 40 V
2 mV
± 100 V
5 mV
a. Accuracy: ± (% of reading or setting value + offset value). Iout is the output current in amps.
b. Vout is the output voltage in volts.
Specifications
Hardware
MPSMU Voltage Range, Resolution, and Accuracy
a
Set Accuracy
± (0.03 % + 900 μV
+ 0.3 × Iout)
± (0.03 % + 4 mV
+ 0.3 × Iout)
± (0.03 % + 7 mV
+ 0.3 × Iout)
± (0.04 % + 15 mV
+ 0.3 × Iout)
VSU (Voltage Source Unit) Specifications
Same as the 4156C VSU.
VMU (Voltage Monitor Unit) Specifications
Same as the 4156C VMU.
8-12
Measurement
Meas.
Reso.
Accuracy
2 μV
± (0.02 % + 700 μV
+ 0.3 × Iout)
20 μV
± (0.02 % + 2 mV
+ 0.3 × Iout)
40 μV
± (0.02 % + 3 mV
+ 0.3 × Iout)
100 μV
± (0.03 % + 5 mV
+ 0.3 × Iout)
Agilent 4155C/4156C User's Guide Vol.1, Edition 11
Max. Current
a
100 mA
100 mA
100 mA (Vout ≤ 20 V)
50 mA
(20 V < Vout ≤ 40 V)
100 mA (Vout ≤ 20 V)
50 mA
(20 V < Vout ≤ 40 V)
20 mA
(40 V < Vout ≤ 100 V)
b