Specifications
This chapter lists complete specifications and supplemental information of Agilent
4155C/4156C and Agilent 41501B. The specifications are the performance
standards or limits against which the 4155C/4156C and 41501B are tested. When
the 4155C/4156C and 41501B is shipped from the factory, it meets the
specifications. The supplemental information are not specifications but are typical
characteristics included as additional information for the operator.
Agilent 4155C/4156C functions:
•
Set measurement and/or stress conditions
•
Control measurement and/or stress execution
•
Perform arithmetic calculations
•
Display measured and calculated results on the LCD display
•
Perform graphical analysis
•
Store and recall measurement setups, and measurement and graphical display
data
•
Dump to printers or plotters for hardcopy output
•
Perform measurement and analysis with internal Instrument BASIC
•
Self-test, Auto-calibration
Agilent 4155C configuration:
•
four medium power source monitor units
(MPSMUs: 10 fA/2 μV to 100 mA/100 V)
•
two voltage source units (VSUs)
•
two voltage monitor units (VMUs)
Agilent 4156C configuration:
•
four high resolution source monitor units
(HRSMUs: 1 fA/2 μV to 100 mA/100 V)
•
two voltage source units (VSUs)
•
two voltage monitor units (VMUs)
Agilent 41501B (Option) configuration:
•
two pulse generator units (PGUs) (Option)
one high power source monitor unit (HPSMU: 10 fA/2 μV to 1 A/200 V)
•
or two MPSMUs (Option).
•
one ground unit (GNDU)
8-2